PN: R109P45
Categories: Medical Oxygen Sensors, Oxygen Sensors (FiO2)
Cross Reference
- Analytical Industries: PSR-11-917-M
- Bird: Avea
- CareFusion: Viasys Avea 16448
- Ceramatec: CAG-12
- City Technology: MOX-9, R-24
- Covidien; Puritan Bennett: 4-074925-00, NP 840
- Datascope: 0600-00-0002 (MAX-12f)
- Draeger: Infinity Delta
- Emerson: Infant Flow
- EnviteC: OOM-202
- GE (Datex-Ohmeda): 6051-0000-222
- Hamilton: 396008, 396009, Galileo, Raphael, Arabella, Aladdin, 51254, G5, Arabella, HM-12, Multinex, Viasys
- Hudson RCI: 5566, V-12
- IMT: PF300
- Infrasonics: Adult Star 601T
- Integral Process: 65016
- International Technologies: M-04
- Maxtec: Max-12 (R109P45)
- MSA: 472062, 470644
- Mercury Medical: 10-103-06
- Nuova: E-17/M
- PB: 760, 840
- Sensoronics; Pacifitech: PT-12A, SS-12A
- Sensidyne: Acutonic Hi Freqency Jet Ventilator, 00M102, SV-12A
- SensorTech: ST-01
- Siemens: 9004979-E347E, 6419332-E380E, 9000, C/D, SV300, SV900, Servo 300, Servo 300A, Servo 700, Servo 710, Servo 900C, Servo 900D, Servo 900E, V710
- Spacelabs: 90513, Capnograph
- Taema: YR049700, Cesar-2, Horus-4
- Teledyne: R24MED, #B-62321, TED 160, TED 190
- Teleflex (Hudson): 5556, V-12
- Ventrex: V-12A
- Viamed: R-24Med
- VTI: 103100
Specifications
- Range of Measurement: 0 to 100%
- Output Range: 12.5 to 16.5 mV
- Zero Offset: ≤ 0.250 mV
- Response Time: ≤ 12 s
- Linearity: ≤ 3%
- Expected Lifetime: > 1,000,000 %O2 hours
- Temperature Compensation: NTC
- Operational Temperature Range: 10˚ to 40˚C
- Storage Temperature Range: -20˚ to 50˚C
- Humidity Range: 0% to 90% RH, non-condensing
- Ambient Pressure Range: 700-1250 mbar
- Interference: PER EN ISO 80601-2-55
- Electrical Interface: 3-pin Molex